Куда я попал?
SECURITM это SGRC система, ? автоматизирующая процессы в службах информационной безопасности. SECURITM помогает построить и управлять ИСПДн, КИИ, ГИС, СМИБ/СУИБ, банковскими системами защиты.
А еще SECURITM это место для обмена опытом и наработками для служб безопасности.

CVE-2018-4846

PUBLISHED 17.09.2024

CNA: siemens

Обновлено: 26.06.2018
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

CWE

Идентификатор Описание
CWE-798 Use of hard-coded credentials

Доп. Информация

Product Status

RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems
Product: RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems
Vendor: Siemens AG
Default status: Не определен
Версии:
Затронутые версии Статус
Наблюдалось в версии RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products affected
 

Ссылки

CVE Program Container

Обновлено: 05.08.2024
SSVC and KEV, plus CVSS and CWE if not provided by the CNA.

Ссылки

Мы используем cookie-файлы, чтобы получить статистику, которая помогает нам улучшить сервис для вас с целью персонализации сервисов и предложений. Вы может прочитать подробнее о cookie-файлах или изменить настройки браузера. Продолжая пользоваться сайтом, вы даёте согласие на использование ваших cookie-файлов и соглашаетесь с Политикой обработки персональных данных.